Criticare
Criticare eQuality 506DN Vital Signs Monitor
Request A Quote
Warranty Information
Files
Click here to find this products equipment manuals
Criticare eQuality 506DN Vital Signs Monitor
Criticare’s eQuality 506DN Vital Signs Monitor is an affordable unit with features that are optimal for both spot-checks and continuous monitoring. Being compact and lightweight, the 506DN is designed for simplicity of use and quick attachment to a roll stand. This unique unit adapts easily to hospital, clinic, and transport requirements.
Includes a 2-year parts and labor warranty.
Features:
- Rechargeable battery with direct AC power
- DOX pulse oximetry with waveform display
- On-screen menu selection for ease of setup and use
- ComfortCuff NIBP measures on-inflation for quicker, more relaxed readings
- Audible and visual alarms are variable for use in continuous monitoring
- Large responsive keys for quick spot-checks with a simple to clean membrane keypad
- Trend memory for blood pressure measurements with heart rate and SpO2 readings
- Accommodates all patient types from adult to neonate
- Nurse call and connectivity to network systems
- Optional integrated Bluetooth
Specifications:
NIBP:
- Technique: oscillometric measure upon inflation
- Measurement Cycle: <40 sec. typical; 0 to 300 mmHg; standard adult cu
- Automatic Measurement: 1, 2, 3, 5, 10, 15, 30, 45, 60, 2-hour, 4-hour and STAT measurements
- STAT Mode: 5 minutes of continuous readings
- Inflation Pressure Range:
- Adult: 30 to 300 mmHg
- Pediatric: 30 to 150 mmHg
- Neonate: 20 to 150 mmHg
- Resolution: 1 mmHg
- Transducer Accuracy: ±2 mmHg or ±2% of reading, whichever is greater over
- Clinical Accuracy: complies with AAMI SP-10:2002
Pulse Oximetry:
- Range: 1% to 99%
- Resolution: 1%
- Accuracy:
- 70% to 99% range: ±2%
- 50% to 69% range: ±3%
- <49%: unspecified
- Method: Dual wavelength LED
- Audio: Tone with each detected pulse; pitch varies with saturation
Heart Rate:
- Range: 30 to 240 bpm
- Accuracy: ±1 bpm or ±1% of reading, whichever is greater; unspecified for <30 or >240 bpm
- Resolution: 1 bpm
- Pulse Volume: 10 levels or Off
Alarms:
- Indication: Audible; Visual
- Volume: 10 levels or OFF
- Silence: Yes; 2 minutes or permanent
Trends:
- Memory: Last 250 NIBP measurements
- Parameters: Systolic, Diastolic, MAP, pulse rate (records SpO2, pulse rate and SpO2 saturation if oximetry is monitored)
Display:
- Type: Monochrome LCD
- Parameters: NIBP, SpO2, pulse rate, plethsymograph
- Status: Alarm Silence; Sensor; Low Battery; External Power; Patient Size
- Languages: English, Danish, Dutch, French, German, Portuguese, Spanish
System Outputs:
- Nurse Call: Contact swtich; 3.5mm phone jack
- COM Port: Serial, RS232 compatible, Bluetooth
- Settings: External printer or CUSP
Physical Properties:
- Weight: 4.5 lbs. (2.0 kgs.)
- Size: 8.0 in (H) x 5.5 in (W) x 5.75 in (D)
Power Requirements:
- Voltage: 100 to 240 VAC; 50/60 Hz
- Battery: 1 sealed lead acid
- Battery Life: 8.0 hours, typical
- Recharge Time: 6.0 hours, typical
Classification:
- Electrical Protection: Class I Equipment
- Degree of Protection: Type CF, Defibrillator-proof
- Protection Against INgress: IPX1, Drip-proof Equipment