Description
Criticare eQuality 506DN Vital Signs Monitor
Criticare’s eQuality 506DN Vital Signs Monitor is an affordable unit with features that are optimal for both spot-checks and continuous monitoring. Being compact and lightweight, the 506DN is designed for simplicity of use and quick attachment to a roll stand. This unique unit adapts easily to hospital, clinic, and transport requirements.
Includes a 2-year parts and labor warranty.
Features:
- Rechargeable battery with direct AC power
- DOX pulse oximetry with waveform display
- On-screen menu selection for ease of setup and use
- ComfortCuff NIBP measures on-inflation for quicker, more relaxed readings
- Audible and visual alarms are variable for use in continuous monitoring
- Large responsive keys for quick spot-checks with a simple to clean membrane keypad
- Trend memory for blood pressure measurements with heart rate and SpO2 readings
- Accommodates all patient types from adult to neonate
- Nurse call and connectivity to network systems
- Optional integrated Bluetooth
Specifications:
NIBP:
- Technique: oscillometric measure upon inflation
- Measurement Cycle: <40 sec. typical; 0 to 300 mmHg; standard adult cu
- Automatic Measurement: 1, 2, 3, 5, 10, 15, 30, 45, 60, 2-hour, 4-hour and STAT measurements
- STAT Mode: 5 minutes of continuous readings
- Inflation Pressure Range:
- Adult: 30 to 300 mmHg
- Pediatric: 30 to 150 mmHg
- Neonate: 20 to 150 mmHg
- Resolution: 1 mmHg
- Transducer Accuracy: ±2 mmHg or ±2% of reading, whichever is greater over
- Clinical Accuracy: complies with AAMI SP-10:2002
Pulse Oximetry:
- Range: 1% to 99%
- Resolution: 1%
- Accuracy:
- 70% to 99% range: ±2%
- 50% to 69% range: ±3%
- <49%: unspecified
- Method: Dual wavelength LED
- Audio: Tone with each detected pulse; pitch varies with saturation
Heart Rate:
- Range: 30 to 240 bpm
- Accuracy: ±1 bpm or ±1% of reading, whichever is greater; unspecified for <30 or >240 bpm
- Resolution: 1 bpm
- Pulse Volume: 10 levels or Off
Alarms:
- Indication: Audible; Visual
- Volume: 10 levels or OFF
- Silence: Yes; 2 minutes or permanent
Trends:
- Memory: Last 250 NIBP measurements
- Parameters: Systolic, Diastolic, MAP, pulse rate (records SpO2, pulse rate and SpO2 saturation if oximetry is monitored)
Display:
- Type: Monochrome LCD
- Parameters: NIBP, SpO2, pulse rate, plethsymograph
- Status: Alarm Silence; Sensor; Low Battery; External Power; Patient Size
- Languages: English, Danish, Dutch, French, German, Portuguese, Spanish
System Outputs:
- Nurse Call: Contact swtich; 3.5mm phone jack
- COM Port: Serial, RS232 compatible, Bluetooth
- Settings: External printer or CUSP
Physical Properties:
- Weight: 4.5 lbs. (2.0 kgs.)
- Size: 8.0 in (H) x 5.5 in (W) x 5.75 in (D)
Power Requirements:
- Voltage: 100 to 240 VAC; 50/60 Hz
- Battery: 1 sealed lead acid
- Battery Life: 8.0 hours, typical
- Recharge Time: 6.0 hours, typical
Classification:
- Electrical Protection: Class I Equipment
- Degree of Protection: Type CF, Defibrillator-proof
- Protection Against INgress: IPX1, Drip-proof Equipment